title: "Using fluxible with the Li7500"
output: rmarkdown::html_vignette
vignette: >
  %\VignetteIndexEntry{Using fluxible with the Li7500}
  %\VignetteEngine{knitr::rmarkdown}
  %\VignetteEncoding{UTF-8}
bibliography: biblio_phd_zot.bib
csl: emerald-harvard.csl
---

```{r, include = FALSE}
knitr::opts_chunk$set(
  collapse = TRUE,
  comment = "#>"
)
options(tidyverse.quiet = TRUE)
```

This vignette shows how to process ecosystem fluxes (CO~2~ and H~2~O) measured with the LI7500 tent setup presented in @halbritterPlantTraitVegetation2024.
Because of field constraints, this setup differs from the smaller chamber in several points:

- one file per flux
- start and end of files do not represent start and end of measurements because setting up the tent can take more or less time
- for similar reasons, the length of the measurements is not constant

![Tent setup](tent_peru.jpg)


# Importing the files

We import the data with `licoread::import7500`.
```{r reading, message=FALSE}
library(licoread)
li7500_data <- import7500(
  "ex_data/li7500",
  plotinfo = c("site", "treatment", "date", "plot_id", "trial")
)
```

<!-- The `licoread::import7500` function provides `f_fluxid`, `f_start` and `f_end`, meaning we could skip `fluxible::flux_match`. -->
The `licoread::import7500` function provides `f_start` and `f_end`, meaning we could skip `fluxible::flux_match`.
However, `f_start` and `f_end` are based on the start and end of the files, which would require a strict data collection routine.
If such a strict routine did not happen, there are two options to deal with that.

## Making a new field record

If `f_start` and `f_end` cannot be used as start and end of measurement, we can recreate a record file (for example in a spreadsheet) with the real start and end datetime:

- use start and end from `licoread::import7500`
- process fluxes until visualization (`flux_fitting`, `flux_quality` and `flux_plot`)
- based on the plots, create a `record` file with metadata identifying the fluxes (`filename`), start and end (note that `flux_match` requires datetime format)
- rename the columns `f_start`, `f_end` and `f_fluxid` to avoid losing those informations (`flux_match` will overwrite them)
- include `flux_match` before `flux_fitting` in your workflow, and match the fluxes with the record file you created
- adapt the record file if necessary and rerun until `flux_plot`
- once satisfied with cutting and fitting, run `flux_calc` as normally

That makes the processing non homogeneous but still reproducible.

## One side is reliable


If the start or the end are consistent but not the other, you can change the direction of the cutting in `flux_fitting`.
By default, `flux_fitting` cuts the focus window as `start + start_cut` to `end - end_cut`.
The argument `cut_direction` allows to change that:

- `cut_direction = "from_start"` means the focus window will be from `start + start_cut` to `start + end_cut`
- `cut_direction = "from_end"` means the focus window will be from `end - start_cut` to `end - end_cut`


# Processing with `fluxible`
<!-- Since the `licoread::import7500` already provides `f_fluxid`, `f_start` and `f_end`, we can skip `fluxible::flux_match` and directly fit with `fluxible::flux_fitting`. -->


## CO~2~

Wet air correction:
```{r wetair_co2}
library(fluxible)
li7500_data_co2 <- flux_drygas(li7500_data, `CO2 umol/mol`, `H2O mmol/mol`)
```


Fitting a linear model:
```{r fitting_co2, message = FALSE, warning = FALSE}
li7500_fits_co2 <- flux_fitting(li7500_data_co2,
                                f_conc = `CO2 umol/mol_dry`,
                                fit_type = "linear")
```

Using `fluxible::flux_quality` to assess the quality of the dataset.
```{r flags_co2}
li7500_flags_co2 <- flux_quality(li7500_fits_co2,
                                 f_conc = `CO2 umol/mol_dry`,
                                 rsquared_threshold = 0.5)
```


Plotting:
```{r plotting_co2, message=FALSE, warning=FALSE, fig.cap="Output of `flux_plot` for CO~2~ fluxes"}
flux_plot(li7500_flags_co2,
          f_conc = `CO2 umol/mol_dry`,
          f_ylim_upper = 500,
          y_text_position = 450,
          print_plot = TRUE)
```

Can be exported in pdf directly (much easier to handle with a lot of data) with:
```{r plotting_exp_co2_pdf, eval = FALSE}
flux_plot(li7500_flags_co2,
          f_conc = `CO2 umol/mol_dry`,
          f_ylim_upper = 500,
          y_text_position = 450,
          print_plot = FALSE,
          output = "longpdf",
          f_plotname = "li7500_co2")
```

If other data (PAR and co) need processing, that would happen here.
To "compress" environmental variables at a single value per flux, see the arguments `cols_ave`, `cols_sum`, and `cols_med` in `fluxible::flux_calc`.
To keep the raw data (gas concentration or anything else) in a nested column, see the `cols_nest` argument.

Now let's calculate the fluxes with `fluxible::flux_calc`.
```{r calc_co2}
li7500_fluxes_co2 <- flux_calc(li7500_flags_co2,
                               slope_col = f_slope_corr,
                               temp_air_col = Temperature,
                               setup_volume = 2197,
                               atm_pressure = pressure_atm,
                               plot_area = 1.44,
                               conc_unit = "ppm",
                               flux_unit = "umol/m2/s",
                               cols_keep = c(
                                 "site", "treatment", "date", "plot_id", "trial"
                               ))
```

## H~2~O


Wet air correction:
```{r wetair_h2o}
li7500_data_h2o <- flux_drygas(li7500_data, `H2O mmol/mol`, `H2O mmol/mol`)
```


Fitting a linear model:
```{r fitting_h2o, message = FALSE, warning = FALSE}
li7500_fits_h2o <- flux_fitting(li7500_data_h2o,
                                f_conc = `H2O mmol/mol_dry`,
                                fit_type = "linear",
                                start_cut = 0,
                                end_cut = 0)
```

Using `fluxible::flux_quality` to assess the quality of the dataset:
```{r flags_h2o}
li7500_flags_h2o <- flux_quality(li7500_fits_h2o,
                                 f_conc = `H2O mmol/mol_dry`,
                                 rsquared_threshold = 0.5,
                                 ambient_conc = 10, # the default is for CO2
                                 error = 2)
```


Plotting:
```{r plotting_h2o, message=FALSE, warning=FALSE, fig.cap="Output of `flux_plot` for H~2~O fluxes"}

flux_plot(li7500_flags_h2o,
          f_conc = `H2O mmol/mol_dry`,
          print_plot = TRUE,
          f_ylim_lower = 5,
          f_ylim_upper = 15,
          y_text_position = 12)
```

Or directly exported as a pdf:
```{r plotting_h2o_pdf, eval = FALSE}
flux_plot(li7500_flags_h2o,
          f_conc = `H2O mmol/mol_dry`,
          print_plot = FALSE,
          output = "longpdf",
          f_plotname = "li7500_h2o",
          f_ylim_lower = 5,
          f_ylim_upper = 15,
          y_text_position = 12)
```


Flux calculations:
```{r calc_h2o}
li7500_fluxes_h2o <- flux_calc(li7500_flags_h2o,
                               slope_col = f_slope_corr,
                               temp_air_col = Temperature,
                               setup_volume = 2197,
                               atm_pressure = pressure_atm,
                               plot_area = 1.44,
                               conc_unit = "mmol/mol",
                               flux_unit = "mmol/m2/s",
                               cols_keep = c(
                                 "site", "treatment", "date", "plot_id", "trial"
                               ))
```


# Calculating Gross Primary Production and Transpiration

I'm glad you asked!
The function `fluxible::flux_diff` is precisely made for that.

```{r gpp, warning=FALSE}
li7500_fluxes_co2 <- flux_diff(li7500_fluxes_co2, type_col = trial,
                               id_cols = c(
                                 "site", "treatment", "date", "plot_id"
                               ), type_a = "p", type_b = "r",
                               diff_name = "GPP", cols_keep = "all")
```

```{r trans, warning=FALSE}
li7500_fluxes_h2o <- flux_diff(li7500_fluxes_h2o, type_col = trial,
                               id_cols = c(
                                 "site", "treatment", "date", "plot_id"
                               ), type_a = "p", type_b = "r",
                               diff_name = "T", cols_keep = "all")
```

# Further calculations

In order to calculate carbon and water use efficiencies, we need both H~2~O and CO~2~ fluxes in the same dataset.

```{r joining}
library(tidyverse)
# to avoid confusion, we add a gas column
# this might be implemented in flux_calc in the future

li7500_fluxes_co2 <- li7500_fluxes_co2 |>
  mutate(gas = "co2")

li7500_fluxes_h2o <- li7500_fluxes_h2o |>
  mutate(gas = "h2o")

li7500_fluxes <- bind_rows(li7500_fluxes_co2, li7500_fluxes_h2o)

# Now the data are in a single long df
```
```{r, str, message=FALSE, echo=FALSE}

str(li7500_fluxes, width = 70, strict.width = "cut")
```



### References